User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Routine Periodic Maintenance
UM-20006520DE, Rev. A Page 57
Routine Periodic Maintenance
Regular cleaning, inspection, testing, and repair are the basis of any
effective preventive maintenance program. Such a program helps keep the
system in top working order and ensures the reliability of safety interlocks
and fail-safe mechanisms.
A recommended routine inspection and maintenance schedule is provided
below.
Inspection/Service Frequency Performed By Remarks
Routine exterior
cleaning.
As required by hospital/
clinic protocol.
Hospital/Clinic
Staff
None
Inspect cables and all
external surfaces for
damage.
Weekly
Hospital/Clinic
Staff
If damage is found, call Lumenis
Customer Service.
Inspect electrical
connections.
Weekly
Hospital/Clinic
Staff
If damage is found, call Lumenis
Customer Service.
Check remote interlock
connection and
emergency stop button.
Weekly
Hospital/Clinic
Staff
If interlock and/or button do not
perform as required, call
Lumenis Customer Service.
Inspect/replace the
debris shield
Weekly or if required due
to low output energy.
Hospital/Clinic
Staff
If output energy is still low after
replacing the shield, call
Lumenis Customer Service.
Deionizer and particle
filters replacement.
Annually
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.
Electrical safety checks.
Annually (or as required
by institutional
procedures).
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.
Check and perform
energy detectors
calibration procedure.
Annually, or as required if
system does not perform
to specifications, or
occurrence of error
messages.
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.