User's Manual

Table Of Contents
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Message Appears: Check footswitch
UM-20006520DE, Rev. A Page 55
Message Appears: Check footswitch
You are pressing either one both footswitch pedals in Standby mode.
Release the footswitch, switch to Ready and press one of the
footswitch pedals.
Message Appears: Check interlock
The interlock door is open, or the interlock plug is not properly inserted.
Close the interlock door, or insert the interlock plug.
Message Appears: Insert debris shield
The debris shield is missing or is not properly inserted.
Insert the debris shield.
Message Appears: No lasers
The system is experiencing a laser malfunction.
Contact Lumenis Customer Service.
Message Appears: Energy high
The energy delivered is more than 50% higher than the selected level.
•Press the Standby button to clear the message, then press the Ready
button and resume normal operation. If the condition continues, turn
off the laser system for five seconds, then turn it back on. If the
condition persists, contact Lumenis Customer Service.
Message Appears: Energy low
The energy delivered is less than 50% of the selected level.
•Press the Standby button to clear the message, then press the Ready
button and resume normal operation. If the condition continues, turn
off the laser system for five seconds, then turn it back on. If the
condition persists, contact Lumenis Customer Service.