User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems “Popping” or “Tapping” Coming Sound
UM-20006520DE, Rev. A Page 53
“Popping” or “Tapping” Coming Sound from the Fiber Port
This is probably due to a malfunction of the optical fiber connector.
• Replace both the optical fiber and the debris shield.
Fiber Burn Back
Optical fiber burn back may occur during prolonged procedures,
especially when using higher power.
• Renew the optical fiber tip by stripping and cleaving the fiber.
Unrecognized Fiber
1. Replace the optical fiber with a Lumenis compatible one and resume
normal operation.
2. If problem persists, contact Lumenis Service.
A Notification or Error Message Appears on the Control Panel
There are three types of clearable errors:
• An error message that clears automatically (i.e., “Attach an
Authorized Fiber”) when an authorized fiber is attached.
• An error message that clears when the stated problem is resolved
(i.e., “High temperature” will clear when the temperature of the
system goes down).
• An error that requires user acknowledgement after resolving the
problem (i.e., “Energy High”); press the Standby button on the
screen to clear the message.