User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Chapter 7 - Troubleshooting and Mainte-
UM-20006520DE, Rev. A Page 51
Chapter 7 - Troubleshooting and
Maintenance
Handling Error Messages and Notifications
Notifications and error messages appear in the Notification bar at the
bottom of the screen.
1. Follow the instructions in the notification bar if any are given.
2. If instructions appear, perform the required task as detailed in the
message. If the error is resolved, the message will fade and normal
operation may be resumed.
3. Several types of errors will not fade - in a case like this press the
Standby button to clear the notifications or error message; normal
operation may be resumed.
4.
Several types of errors will only appear with a number (“
Fault XXX
”);
press the Standby button to clear the error and try to resume normal
operation. If this does not resolve the problem restart the system.
5. If the problem does not resolve, contact Lumenis Customer Service.
Figure 12: Notification Bar
Fault XXX
Clear kJ