User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Intra-Operative Instructions
UM-20006520DE, Rev. A Page 48
5. Press the Ready button to switch to Ready mode.
WARNING:
Always check your parameter settings on the screen before setting the
system to Ready mode.
6. Verify that your foot is on the appropriate footswitch pedal for the left-
side or right-side parameter settings on the screen.
7. Press the footswitch that corresponds to the desired set of parameters
to deliver the treatment beam.
As the laser delivers the treatment beam, the highlighted yellow
margin appears around the appropriate footswitch settings area on the
control screen and an audible signal sounds to alert you that laser
energy is being emitted. The audible signal is different for the right
and left pedals.
8. Press the Ready and Standby buttons on the screen or use the
footpedal Ready/Standby button (on the top of the footswitch) to
toggle between Ready and Standby modes.
9. If surgery is interrupted, set the laser system to Standby mode to
disable the footswitch.
WARNING:
Always set the laser system to Standby mode when it is not in use to
avoid unintended laser emission.