User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Treatment Screen Description
UM-20006520DE, Rev. A Page 44
Treatment Screen Description
Once the laser self-test and warm-up have successfully completed, the
main Treatment screen displays on the control panel.
The elements of the Treatment screen are detailed as follows (the
numbered arrows in figure 9 correlate to the numbered steps below):
1. Left Pedal Designation – displays which footswitch pedal should be
used for the group of output parameters in the box beneath the
designator.
2. Right Pedal Designation – displays which footswitch pedal should be
used for the group of output parameters in the box beneath the
designator.
3. Treatment Settings for each pedal – each side of the screen defines
the Energy, Frequency and Pulse width settings for lasing when the
corresponding pedal is pressed. Press the ► or ◄ selectors to reach
the desire settings; the derived power will be displayed
4. Pulse Width – the user may set the laser emission mode to short or
long pulses. Toggle this button to select Long or Short pulses.
5. Aiming Beam – the Aim Beam display shows the selected aiming
beam intensity: high, medium, low or off.
• At laser system turn on the aiming beam setting retains the intensity
setting from the last use of the laser system, or it will automatically
select the minimum setting it the aiming beam was turned off at the
last shut-down.
•Press the Aim Beam ► or ◄ selectors to set the aiming beam
intensity.
• To turn off the aiming beam, press the ◄ selector until no bars are
illuminated.
CAUTION:
If the aiming beam is turned off, ensure that the tip of the optical fiber and
the targeted surgical site are both under direct visualization.