User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Safety Eyewear
UM-20006520DE, Rev. A Page 43
Safety Eyewear
Verify that all persons in the operating room are equipped with
appropriate laser safety eyewear.
Verification of Connections
1. Verify that the footswitch is properly connected.
2. Verify that the remote interlock connector is connected.
Powering on the System
1. Set the laser system’s main power circuit breaker to the On (up)
position.
2. Insert the key into the keyswitch (Refer to figure 8.) and rotate the key
to the || (start) position; hold for one full second and release the key.
Upon release, the spring-loaded key rotates to the |(on) position.
A laser self-test and warm-up begin. The self-test and warm-up take
approximately one minute. As internal tests are performed, self-test
pass/fail messages display on the control screen. When the self-test is
successfully completed, the main Treatment screen displays on the
control panel.
NOTE:
If any fault conditions are encountered during laser system start-up and
self-test, error messages can appear in the notification area on the control
screen. Refer to the Troubleshooting section later in this manual.
Restarting the Laser System
If it becomes necessary to restart the system:
1. Turn the keyswitch to the (off) position.
2. Wait 5 seconds, then rotate the keyswitch to the || (start) position; hold
for one full second, and release the key.