User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Connecting the External Door Interlock
UM-20006520DE, Rev. A Page 36
Connecting the External Door Interlock Connector
The external door interlock is a safety feature that disables the laser
system if the treatment room doors are opened or the external door
interlock connector is removed while the laser system is in Ready mode.
The laser system remains inoperative until the connector is inserted.
1. Align the pins of the external door interlock connector with the socket
of the external interlock receptacle.
2. Insert the external interlock connector into the external interlock
receptacle.
3. Turn the metal lock clockwise until it screws in.
4. If the treatment door is opened or if the external door interlock
connector is removed, the laser system automatically disables and
returns to Standby mode and a notification appears in the notification
bar.
5. To resume treatment, close the treatment room door or reinsert the
external door interlock connector, and press the Ready button.
Figure 5: Connect the External Door Interlock Connector