Product Info

Report No: CCISE191202301-V01
Shenzhen Zhongjian Nanfang Testing Co., Ltd. Project No.: CCISE1912023
No. B-C, 1/F., Building 2, Laodong No.2 Industrial Park, Xixiang Road,
Bao’an District, Shenzhen, Guangdong, China
Telephone: +86 (0) 755 23118282 Fax: +86 (0) 755 23116366, E-mail: info@ccis-cb.com Page 71 of 171
Test Laboratory: CCIS Date/Time: 12.17.2019 08:08:42
Communication System: UID 0, CW (0); Frequency: 835 MHz; Duty Cycle: 1:1
Medium parameters used: f = 835 MHz; σ = 0.916 S/m; ε
r
= 41.138; ρ = 1000 kg/m
3
Phantom section: Flat Section
DASY5 Configuration:
Probe: EX3DV4 - SN3924; ConvF(9.67, 9.67, 9.67); Calibrated: 08.30.2019;
Sensor-Surface: 1.4mm (Mechanical Surface Detection)
Electronics: DAE4 Sn1373; Calibrated: 08.09.2019
Phantom: SAM 5.0; Type: QD000P40CD; Serial: TP:1765
Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)
System Performance Check at Frequency 835 MHz Head Tissue/d=15mm,
Pin=80 mW, dist=2.0mm (EX-Probe)/Area Scan (41x131x1): Interpolated grid:
dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 1.09 W/kg
System Performance Check at Frequency 835 MHz Head Tissue/d=15mm,
Pin=80 mW, dist=2.0mm (EX-Probe)/Zoom Scan (7x7x7) (5x5x7)/Cube 0:
Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 32.08 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 1.18 W/kg
SAR(1 g) = 0.776 W/kg; SAR(10 g) = 0.519 W/kg
Maximum value of SAR (measured) = 1.03 W/kg
0 dB = 1.03 W/kg = 0.13 dBW/kg