The dielectric parameters of liquids were verified prior to the SAR evaluation using a Speag Dielectric Probe Kit
and an Agilent Network Analyzer.
The following table shows the measuring results for simulating liquid.
Frequency
(MHz)
Liquid
Type
Liquid
Temp.
()
Conductivity
(σ)
Permittivity
(εr)
Conductivity
Target(σ)
Permittivity
Target(εr)
Delta
(σ)%
Delta
(εr)%
Limit
(%)
Date
(mm/dd/yy)
835
Head
22.4
0.92
41.14
0.9
41.5
2.22
-0.87
±5
12.17.2019
1800
Head
22.8
1.38
40.18
1.40
40.0
-1.43
0.45
±5
12.11.2019
1900
Head
22.8
1.41
39.78
1.4
40.0
0.71
-0.55
±5
12.11.2019
2450
Head
22.7
1.82
39.48
1.8
39.2
1.11
0.71
±5
12.10.2019
2600
Head
22.7
2.01
38.53
1.96
39.0
2.55
-1.21
±5
12.10.2019
835
Body
22.3
0.99
54.68
0.97
55.2
2.06
-0.94
±5
12.13.2019
1800
Body
22.4
1.50
52.82
1.52
53.3
-1.32
-0.9
±5
12.09.2019
1900
Body
22.4
1.54
52.31
1.52
53.3
1.32
-1.86
±5
12.09.2019
2450
Body
22.8
1.95
53.29
1.95
52.7
0
1.12
±5
12.23.2019
2600
Body
22.8
2.19
52.07
2.16
52.5
1.39
-0.82
±5
12.23.2019