Fig. 9.5 Photo of Liquid Height for Head SAR
(2000MHz~2600MHz) (depth>15cm)
Fig. 9.6 Photo of Liquid Height for Body SAR of Twin
Phantom (2000MHz~2600MHz) (depth>15cm)
The relative permittivity and conductivity of the tissue material should be within ±5% of the values given in the
table below recommended by the FCC OET 65 supplement C and RSS 102 Issue 5.
Target Frequency
Head
Body
(MHz)
εr
σ(S/m)
εr
σ(S/m)
150
52.3
0.76
61.9
0.80
300
45.3
0.87
58.2
0.92
450
43.5
0.87
56.7
0.94
835
41.5
0.90
55.2
0.97
900
41.5
0.97
55.0
1.05
915
41.5
0.98
55.0
1.06
1450
40.5
1.20
54.0
1.30
1610
40.3
1.29
53.8
1.40
1800-2000
40.0
1.40
53.3
1.52
2450
39.2
1.80
52.7
1.95
3000
38.5
2.40
52.0
2.73
5800
35.3
5.27
48.2
6.00
( εr = relative permittivity, σ = conductivity and ρ = 1000 kg/m