User guide
Sponsler, Inc. Model T575N Temperature Compensated Totalizer/Printer
EXAMPLE 5:
From the flowmeter calibration sheet the full-scale flowrate is 898 cycles
per second.
K-Factor (EXAMPLE 3) = 60.759
Maximum Count =
Full-scale flowrate frequency X 60
K-Factor
Maximum Count =
898 x 60
60.759
Maximum Count =
53880
60.759
Maximum Count = 886.78
Maximum Count is not exceeded; no pulses will go into storage.
Model T575N Option ‘IC’ EPROM
POSITION
PRODUCT
LOW TEMP
REF. TEMP
HI TEMP
0
Oxygen
-186.046
-183.11
-167.790
1
Nitrogen
-198.255
-195.930
-181.952
2
Argon
-187.209
-186.046
-167.790
3
CO2
-40.050
-16.388
-6.929
4
N20
-51.087
-34.679
15.384
5
CACO2
-40.051
-25.8333
-6.969
6
MAPP
-17.871
15.384
50.921
7
LPG
-40.050
15.384
49.935
8 Calibrate &
switch test
9
Self test
Temp = °C
7