Setup & Operation User guide

16
STEP 4: SYSTEM CALIBRATION CONTINUED
T UNIT^
This field allows you to choose the temperature units - Degrees Fahrenheit or Degrees Celsius. Press
ENTER and the cursor will drop to the bottom line. Use
LL
LL
L to scroll between the two choices, DEG. F
and DEG. C. When the desired choice is displayed, press ENTER and the cursor will move to the top
line.
Press
LL
LL
L and the cursor will move to:
RTD SLP
Factory calibration setting only. Press
LL
LL
L and the cursor will move to:
RTD OFS
Factory calibration setting only. Press
LL
LL
L and the cursor will move to:
FLOW DIR^
This field allows you to match the LCR flow direction to the meter's pulse output. If the counter runs
backwards, this arrow should be changed. The choices are
II
II
I or
JJ
JJ
J. Press ENTER and the cursor will
drop to the bottom line. Use
LL
LL
L to scroll between the two choices. When the desired choice is displayed,
press ENTER.
Press
LL
LL
L and the cursor will move to:
LAST CALIBRATED
This field contains the date and time of the last calibration. It cannot be edited.
Press
LL
LL
L and the cursor will move to:
LCR #
This field contains the node address for RS485 communication. Enter a number between 1 and 250.
The default value is 250.
Press
LL
LL
L and the cursor will move to:
CALIB #
For metrology use only. This is the number of times the calibration switch position has been entered.
DISPLAY ONLY.
CALIB EVENT
For metrology use only
. This is the number of times the calibration has been changed. DISPLAY ONLY.
CONFIG EVENT
For metrology use only . This is the number of times the configuration has been changed. DISPLA Y
ONLY.
Press the M1 key to return to the top of the SYSTEM CALIBRATION menu. Then press
LL
LL
L to scroll to PRODUCT &
SHIFT INFORMATION.
Lap Pad Calibration & Set-up
System Calibration - Screen 4
FLOW DIR^
-->
LAST CALIBRATED
7/11/2002 11:44
LCR #
250
System Calibration - Screen 5
CONFIG EVEN
T
1
CALIB EVEN
T
19
CALIB#
34