Setup Owner manual

40
Screen 10c: Product Calibration (3/6)
Linearization Mode is used either to enable multi-point
linearization in the LectroCount, or to disable linearization
so base calibration can be set up. Press ENTER to open
a list box window with the two options:
Setup
Applied
Select Setup for single-point calibration. Select Applied
for multi-point calibration.
NOTE: Always perform initial calibration in Setup.
Single-point and Multi-point Calibration are covered on
Pages 44-47.
Pulses/Unit represents number of pulse edges counted
by the LectroCount per unit of measure. This number
is also known as the product’s K-factor. This number
is used to scale Prover Quantity. Only gross volume
applies. Press ENTER to open a eld edit window and
enter a value based on the following guideline:
Decimal Setting Range
Whole 2.0 to 20000.0
Tenths 20.0 to 20000.0
Hundreths 20.00 to 20000.0
Since this will change during calibration, leave the default
value of 0 and proceed to rest of editable elds in Product
Calibration.
Prover Quantity displays amount of product measured
during the last prover run. When this value is overwritten
with actual Prover Volume read during a calibration run, a
new Pulses/Unit will be recalculated automatically. This
eld should only be edited during meter calibration.
NOTE: If this is the initial calibration for the meter, enter
the rest of the Product Calibration data before entering
this eld.
Press Next (F1) to advance to Screen 10d on page 44.
Product Calibration
(3/6) 1
Units: Gallons
Setup
492.506104
---------
PRODUCT CALIBRATION
Product #: 1
Linearization Mode:
Pulses/Unit:
Prover Quantity:
Next Prev Cnfg Main