User guide






 
 

 

 

 



 
 
 
x10
-1
Value to the right
of decimal point determines
the negative power
x10
-1
= One decimal place to the left
Real value move decimal
point to the left at x10
-x
.
x = number at the far right
Sets decimal point on the
display counter to hundreths

K-factor
K-factor =
(1 - )Approximate K-factor
ExamplE
191.997696 =
(1 -
)192
Approximate K-factor for the CIM100 is
726 pulses per gallons
192 pulses per litre
Approximate C-factor for the CIM100 is
726 = 0.001377410
192 = 0.005208333
c-factor
C-factor =
X
K-factor
ExamplE
0.005208396 =
1
191.997696
% Error
Prover Volume - Metered Volume
Prover Volume
% Error =
100 - 99.88
100
.0012 =
ExamplE

If the K-factor is 100, the
C-factor is ¹⁄₁₀₀ or 0.01.






The series of equations apply the CIM100
totalizer volume and the prover volume
recorded during the proving run in order to
calculate the C-factor.


.0012
100
% Error
100
