Datasheet

LTC2990
17
Rev. F
For more information www.analog.com
APPLICATIONS INFORMATION
Table8. Temperature Measurement MSB Data Register Format
BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 BIT 0
DV* SS** SO
D12 D11 D10 D9 D8
*DATA_VALID is set when a new result is written into the register.
DATA_VALID is cleared when this register is addressed (read) by the I
2
C
interface.
**Sensor Short is high if the voltage measured on V1 is too low
during temperature measurements. This signal is always low for T
INT
measurements.
Sensor Open is high if the voltage measured on V1 is excessive
during temperature measurements. This signal is always low for T
INT
measurements.
Table9. Temperature Measurement LSB Data Register Format
BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 BIT 0
D7 D6 D5 D4 D3 D2 D1 D0
Table6. Voltage/Current Measurement MSB Data Register
Format
BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 BIT 0
DV* Sign D13 D12 D11 D10 D9 D8
*Data Valid is set when a new result is written into the register. Data Valid
is cleared when this register is addressed (read) by the I
2
C interface.
Table7. Voltage/Current Measurement LSB Data Register
Format
BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 BIT 0
D7 D6 D5 D4 D3 D2 D1 D0
Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.Downloaded from Arrow.com.