Specifications

Page 99 of 152
The ACM (Mini Attendant Control) box is fitted to hospital and care beds for use where patient positioning
must be carefully controlled by the medical staff.
The compact design and simple operation makes it easy for the nursing staff to retain direct control over
critical positioning functions whilst giving the patient a limited degree of adjustment.
The ACM must be mounted correctly on a flat surface to ensure IP degree. A short circuit in the cable can
cause movement. To possibly avoid this risk, choose a OpenBus system.
4. ACM (MEDLINE
®
CARELINE
®
)
The ACO (Attendant Control OpenBus
TM
) is a cost optimised and compact unit with up to 15 buttons that can
be used as Handset keys or lock-outs. The lock-out function can be made visable by using yellow LED´s.
The ACO is compatible with control boxes that use an OpenBus interface i.e. CB6S, CB16 and CB20.
Usage:
Compatible with OpenBus™ e.g. CB20, CB6S and CB16.
Ambient temperature: +5˚ to + 40˚C
Approvals: IEC 60601-1 Edition 3.1 (2012)
IEC 60601-1-6:2010 + A1:2013,
ANSI/AAMI ES60601-1: A1:2012, C1:2009/(R) 2012 and A2:2010/(R)2012,
CAN/CSA-C22.2 NO. 60601-1:14
Current consumption: 15 mA (Standard versions with 5 no. LEDs (5 LEDs x 2 mA = 10 mA)
+ microprocessor 5 mA, total = 15 mA
In order to comply with the norm, the ACO must hang vertically from its hook during the washing process.
5. ACO (MEDLINE
®
CARELINE
®
)
Recommendations
Always use Locking ring and cables with O-rings.
Locking ring and cables with O-rings must be fitted to ensure IP degree.
If other front covers than standards are requested, the customer must design them.
N.B.
Cables are inclusive an O-ring.
ACOM is the obvious control for hospital and nursing home beds where
patient positioning needs careful control by medical staff. ACOM is an
OpenBus
TM
control.
Usage:
Operation temperature: +5 °C to + 40 °C
Storage temperature: -10 °C to + 50 °C
Relative humidity: 20 % to 80 % at +30° C, non-condensing
Atmospheric pressure: 700 to 1060 hPa
Approvals: IEC 60601-1 Edition 3.1 (2012), IEC 60601-1-6:2010 + A1:2013
Compatibility: Compatible with LINAK OpenBus control boxes, CO-generation
6. ACOM (MEDLINE
®
CARELINE
®
)
Recommendations
Clean the hand control regularly to ensure good hygiene standards.
When a defective ACOM is replaced, check that the new ACOM has exactly the same specification and functionality.
Do not submerge the hand control in water.
Unless otherwise specified or agreed with LINAK, the hand control is only intended to be used for LINAK systems.
When changing hand controls for OpenBus systems, the power must be switched off.
It is recommended to check the hand control and cable for damage and holes caused by violent handling before washing the application or at
least once a year.
It is recommended to have a parking place for the hand control on the application where the customer ensures that the hand control does not fall off.
Warnings:
Do not sit or lie on the hand control as this can cause unintended movement of the application.