User's Manual

MAINTENANCE AND TROUBLESHOOTING
Calibration Adjustments
24 LDT-5525
CHAPTER 3
ITE Current Calibration
The following procedure is for calibrating the ITE constant current source for both
polarities of current. During this procedure the ITE current is driven to a series of
pre-determined values. When each of these values is reached and is stable, the
user enters the actual value of the current, as measured by an external DMM. The
LDT-5525 Temperature Controller then automatically calibrates the TEC current
source and limits.
Calibration may be aborted by pressing the OUTPUT switch.
a Set the sensor select (back panel) switch to "100 uA." Set C1 to 0.99, C2 to 2.57.
Connect an (approximately) 15 k resistor to the sensor pins (7 and 8). Connect a
1, 20 W, resistor across the TEC output terminals (pins 1 and 3) and use a
calibrated DMM to measure the voltage across the resistor. Calculate the current in
the following steps by using Ohm's Law:
I = E / R
-where E is the accurately measured voltage across the resistor, and R is the
accurately measured load resistance. A 4-point probe resistance measurement is
recommended.
b Press and breifly hold in both the MODE and DISPLAY switches. This will put the
LDT-5525 into ITE calibration mode. Wait for three seconds for the output to settle to
about 3 Amps.
c Press and hold in the ENBL switch and turn the ADJUST knob until the display
shows the correct value (absolute value of the ITE measurement), as calculated from
Step a.
d Release the ENBL switch. Wait three seconds to allow the ITE current to settle at the
new set point.
e Repeat Steps c and d four more times, once for each of the (automatically set) set
points -3 Amps, +3 Amps, +1 Amp, and -1 Amp. After the value for the -1 Amp (last)
set point is entered, the LDT-5525 Temperature Controller will automatically calibrate
its ITE current limits.
After about 10 seconds, the LDT-5525 Temperature Controller will store the new
calibration data in non-volatile memory.
f After calibration, the I LIMIT will be automatically set to 0.00 Amps. Reset the I LIMIT
to the desired value.