User's Manual

INTRODUCTION AND SPECIFICATIONS
Specifications
4 LDT-5525
CHAPTER 1
6 Accuracy figures quoted are typical for a calibrated 10K thermistor; accuracy figures are relative to calibration
standard and are dependent on the user-defined configuration of the instrument; variation from typical value is
largely due to uncertainty in thermistor calibration
Thermistor Range (10 µA) 0.0 to 450.0 K
Thermistor Range (100 µA) 0.0 to 45.0 K
Usable Thermistor Range 25 to 450,000 , typical
Thermistor Resistance Resolution
(10 µA)
0.1 k
Thermistor Resistance Resolution
(100 µA)
0.01 k
Thermistor Resistance Accuracy +0.05% of FS
AD590 Reverse Bias 8 Volts
LM335 Bias 0.6 mA
User Calibration
Thermistor Steinhart-Hart equation (2 constants)
IC Sensor 2-point
Measurement (Display
Display Type 4-digit LED
TE Current Range -4.00 to 4.00 Amps
TE Current Resolution 0.01 Amps
TE Current Accuracy +0.03 Amps
Temperature Range -99.9 to 999.9
o
C
Temperature Resolution 0.1
o
C
Temperature Accuracy
6
0
o
C: 0+0.6
o
C, typical
20
o
C: +0.3
o
C, typical
30
o
C: +0.4
o
C, typical
50
o
C: +0.6
o
C, typical