User's Manual

Page 63 of 72
1000001386A
Operational relative humidity
10 % to 90 % (non-
condensing)
Storage temperature (≤ 30 days)
+0⁰C to +45⁰C (32⁰F to 113⁰F)
Storage temperature (> 30 days)
+5⁰C to +27⁰C
(41⁰F to 80.6⁰F)
Storage relative humidity
30 % to 75 % (non-
condensing)
Storage pressure
700 hPa to 1060 hPa
Ingress protection
IP44
Physical
Characteristics
Dimensions
115 mm x 110 mm
Weight
< 30 gm
Color
White or off white
Wear life
120 hours
Wireless Specifications
Frequency band
2.400-2.4835 GHz
Bandwidth
20 MHz
Transmitter power
0 dBm
Modulation type
Complementary Code Keying (CCK)
and Barker Coding
Data Rate
1, 2, 5.5 and 11 Mbps
Quality of serviceRange
5 meters (Note*)
Wireless security
WPA2-PSK
Declaration and Disclaimers
Other equipment may interfere with the operation
of the Biosensor, even if it complies with CISPR
EMISSION requirements. Portable and mobile RF
Communication equipment can affect Biosensor
operation.
Note*: QoS verified for 10 meters range in bench setup.
**: ECG Signals shall recover within 6 seconds after removal of defibrillation voltage
11.2. Table 2: Platform (LX1550) Heart Rate, RR, Skin temperature and
Accelerometer Specifications