User's Manual

Page 62 of 72
1000001386A
11. Appendix
11.1. Table 1: Biosensor (LP1550) Specifications
Specifications
ECG channels
Two (2) ECG-A and ECG-D
Classification of Applied parts
Defibrillation proof type CF**
Frequency Response
0.2 Hz to 40 Hz
CMRR
> 90dB
Input impedance
> 10 Meg ohms at 10Hz
ADC sampling rate
244.14 SPS
ADC resolution
16 bits
Noise
<30uVpp
Dynamic Range
±300mV
Lead off detection
Injected Current
< 10uA
Injected Frequency
10 KHz
Power
Requirements
Battery Type
Li-MnO2
Battery Life
Biosensor shall have a battery life of minimum
120 hours of normal operation (continuous
transmission) under normal wireless
environment
Battery Capacity
1200 mAh
Output Voltage
3 V
Charging Mode
Not rechargeable
Environmental Specifications
Transportation temperature (≤ 10 days)
-5⁰C to +50⁰C (23F to 122⁰F)
Operational temperature
+0⁰C to +45⁰C (32⁰F to 113⁰F)