Use Instructions

53
Questions? Call 1-800-635-6156
• UL 1637:2017 Standard for Home Health Care Signaling Equipment.
Open source software
The GoSafe 2 Mobile Pendant (7250MHB) utilizes the Wiced™ SDK and associated
third party, open-source code and object les which are used under license from their
respective providers within the SDK. Wiced is a trademark of Cypress Semiconductor
Corporation.
Environmental Ratings
Operating Storage Bathing*
Temperature 32° F to 104° F
(0° C to 40° C)
-4° F to 140° F
(-20° C to 60° C)
Up to 122° F
(50° C)*
Relative
Humidity
10 to 90%
(non-condensing)
10 to 90%
(non-condensing)
10 -90%
(non-condensing)
Atmospheric
Pressure
101 kPa to 80 kPa
(approximately
0-6600 ft/0-2000 m)
N/A 101 kPa to 80 kPa
(approximately
0-6600 ft/0- 2000 m)
Altitude 6,600 feet (2 km)
Maximum
N/A 6,600 feet (2 km)
Maximum
* Up to 30 minutes exposure.
Electrical Specications
GoSafe 2 Mobile Pendant (7250MHB)
AC Voltage Source
[1]
(VAC) N/A - EUT battery operated. AC battery charger
adapter 100-240Vac 50/60Hz, 0.5A
DC Power Source (VDC) 3.7 V, 920mAh
[2]
, Lithium Ion Rechargeable
[3]
Battery
Type of Protection Against
Electric Shock
GoSafe 2 is battery operated device. Protection
against electric shock is provided by Class II
approved medical grade power supply adapter.
Power supply adapter mains plug provides
protection from AC supply mains on all poles.
Degree of Protection
Against Electric Shock
Type BF Applied Part
Degree of Protection
Against Ingress of Water
Water Resistant (1 meter, for 30 Minutes), IPX7
Mode of Operation Continuous Operation
Sound Pressure Level 116-123 dB at 1 kHz (measured 1 cm from the speaker)
MHB3_7250_IFU 2020c.indd 53 2/15/21 2:28 PM