Use Instructions

43
Questions? Call 1-800-635-6156
Software Failure-End Device 7250 MHB
Alarm Event
Description
In the event that an End Device senses it has a
software error, a warning message may be sent to
the Response Center depending on the nature of
the error.
Priority Medium
Device Action May have limited operation, depending on
the failure
Alarm Event
Reported/
Displayed
Locally?
No
Alarm Event
Reported to
Lifeline?
Yes. A silent maintenance signal is sent to the
Response Center if the nature of the error does
not impair signal transmission function. Upon
receiving the maintenance signal, the Response
Center prioritizes the response and contacts the
Subscriber to verify the status of their equipment.
If the End Device is not capable of sending a signal
due to the nature of the error, exception reporting
at the Response Center that does not require a
signal from the End Device will identify the device
as requiring maintenance action.
MHB3_7250_IFU 2020c.indd 43 2/15/21 2:28 PM