Use Instructions
42
Questions? Call 1-800-635-6156
Device Hardware Failure – 7250 MHB
Alarm Event
Description
In the event that internal diagnostics indicate that
the End Device has a Hardware Error, a Warning
Message is displayed, if possible, based on the
nature of the failure.
Priority Medium
Device Action May have limited operation, depending on
the nature of the failure
Alarm Event
Reported/
Displayed
Locally?
No.
Alarm Event
Reported to
Philips Lifeline?
Yes. A silent Maintenance Signal is sent to the
Response Center if the nature of the failure does
not impair the signal transmission function. Upon
receiving the Maintenance Signal, the Response
Center prioritizes the response and contacts the
Subscriber to verify the status of their equipment.
If the End Device is not capable of sending a signal
due to the nature of the error, exception reporting
at the Response Center that does not require a
signal from the End Device will identify the device
as requiring maintenance action.
MHB3_7250_IFU 2020c.indd 42 2/15/21 2:28 PM