Immunity
Test
EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment –
Guidance
Conducted
RF EN/IEC
61000-4-6
3 Vrms 150 kHz
to 80 MHz
6Vrms in ISM
and Amateur
Radio Bands
3 Vrms
6Vrms in ISM
and Amateur
Radio Bands
D=1.17√P
D=0.35√P
80 to 800 MHz
D=0.70√P
800 MHz to 2.5 GHz
D=0.6√P
800 MHz to 2.7GHz
Radiated RF
EN/IEC
61000-4-3
10 V/m
80 MHz to
2.5 GHz
10 V/m
80 MHz - 2.7
GHz
80 % AM 1 kHz
and proximity
elds (7200C)
10V/m
10V/m and
Proximity
Fields
(7200C)
Portable and mobile communications
equipment should be separated from
the HomeSafe/GoSafe by no less
than the distances calculated/listed
below:
D=(3.5/V1)(√P)
D=(3.5/E1)(√P)
80 to 800 MHz
D=(7/Eq)(√P)
800 MHz to 2.5 GHz
D=(6/Eq)(√P) (7200C model)
where P is the maximum power in
watts (W) and D is the recommended
separation distance in meters
(m). Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey,(a) should
be less than the compliance level in
each frequency range.(b) Interference
may occur in the vicinity of
equipment marked with the following
symbol:
NOTE: Conducted RF Immunity does not apply to internal battery operated devices
(7000PHB, 7000PHW, 7000AHB, 7000MHB and 7100MHB). Limit above is set for the
battery charger for 7200C and 7000L communicators.