Use Instructions
92
Questions? Call 1-800-635-6156
Immunity Test EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment –
Guidance
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines EN/
IEC 61000-4-11
(only for devices with
battery charger power
supply-7000MHB
& 7100MHB, 7000L,
7200C)
<5% UT
(>95% dip in U
T
) for
0.5 cycle. 40% U
T
(60% dip in U
T
) for
5 cycles.
70% U
T
(30% dip
in U
T
) for 25 cycles
5% U
T
(>95% dip in
U
T
) for 5 sec.
For 7200C Model:
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle &
70 % UT; 25/30
cycles Single
phase: at 0°
0 % UT; 250/300
cycle (voltage
interruption)
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle. 40%
UT (60% dip in U
T
)
for 5 cycles.
70% U
T
(30% dip
in U
T
) for 25 cycles
5% UT (>95% dip
in U
T
) for 5 sec.
For 7200C Model:
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0 % UT; 1 cycle &
70 % UT; 25/30
cycles Single
phase: at 0°
0 % UT; 250/300
cycle (voltage
interruption)
Electrical power
quality should be
that of a typical
commercial or
hospital environment.
During charging, if the
the user of the GoSafe
(7000MHB/7100MHB)
requires continued
operation during
power mains
interruptions, it is
recommended that
the GoSafe device
power adapter is
powered from an
uninterrupted power
supply or a battery.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Table 3 – Guidance and Manufacturer’s declaration – Immunity
The HomeSafe/GoSafe are intended for use in the electromagnetic environment
specied below. The customer or user of the HomeSafe/GoSafeshould ensure that
it is used in such an environment.