Table 2 – Guidance and Manufacturer’s declaration – Immunity
The HomeSafe/GoSafe devices are intended for use in the electromagnetic
environment specied below. The customer or user of the HomeSafe/GoSafe should
ensure that it is used in such an environment.
Immunity Test EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment –
Guidance
Electrostatic
Discharge (ESD)
EN/IEC 61000-
4-2
±6kV Contact
±8kV Air
±8 kV contact
discharge
±2, 4, 8, 15 kV air
discharge (7200
model)
±6kV Contact
±8kV Air
+/- 8KV Contact
+/- 15KV Air
(7200C model)
Floors should be
wood, concrete or
ceramic tile. If oors
are synthetic the
relative humidity
should be at least
30%
Electrical fast
transient/burst
EN/IEC 61000-
4-4
(only for devices with
battery charger power
supply-7000MHB
& 7100MHB, 7000L,
7200C)
±2 kV for power
supply lines ±1 kV
for input/output
lines
±2 kV for power
supply lines
Electrical power
quality should be
that of a typical
commercial or
hospital environment.
Surge EN/IEC
61000-4-5
(only for devices with
battery charger power
supply- 7000MHB
& 7100MHB, 7000L,
7200C)
±1 kV line(s)
to line(s)
±1 kV line(s)
to line(s)
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power Frequency
50/60Hz
Magnetic Field
EN/IEC 61000-
4-8
3A/m
30A/m
(7200C Model)
3A/m
30A/m
(7200C Model)
Power frequency
magnetic elds
should be that of a
typical commercial or
hospital environment.