SUPERVISION FAILURE (7000PHB/7000PHW/7000AHB/ 7000MHB/7100MHB)
Alarm Event
Description
End Devices periodically check-in to the Communicator.
If the Communicator has not received a check-in from the
End Devices for a week, then a silent alarm is sent to the
Response Center
Priority N/A
Device Action Operates normally
Alarm Event
Reported/Displayed
Locally?
No visual or audible alarm is generated by the devices.
Alarm Event
Reported
to Lifeline?
Yes. A silent Maintenance Signal is sent to the Response
Center by the Communicator. Upon receiving the
Maintenance Signal, the Response Center prioritizes the
response and contacts the Subscriber to verify the status of
their equipment.
ac Power Restored - Communicator (7200C/7000L)
Alarm Event
Description
If ac power is restored within 24 hours of the Communicator’s
“Low Battery + NO ac Power” alarm event occurring, then
the Communicator will report a silent alarm to the Response
Center
Priority N/A
Device Action Operates normally
Alarm Event
Reported/Displayed
Locally?
No visual or audible alarm is generated when ac power
is restored. Instead, the ashing message light / alarm
generated by the “ac power lost” is reset and the Message
button stops ashing.
Alarm Event
Reported
to Lifeline?
Yes. If ac power is restored within 24 hours of a reported low
battery + no ac condition, a Maintenance Signal is sent to
the Response Center, letting the Response Center know the
Communicator is receiving power.