Use Instructions

68
Questions? Call 1-800-635-6156
DEPLETED BATTERY End Device (7000MHB/7100MHB)
Alarm Event
Description
When the charge level of the GoSafe Mobile Button
rechargeable battery is critically low (below 3.7V), a Warning
Message is displayed.
Priority Medium
Device Action Operates normally until the battery is fully depleted.
Alarm Event
Reported/
Displayed Locally?
Yes. GoSafe Mobile Button LED ashes orange, plays single
sound/tone, and a Voice Message. The Message Button on
the Communicator ashes, when the End Device is within
range. Voice Message plays upon pressing the Message
Button. This alarm will self cancel once the battery charge
reaches an acceptable level.
Alarm Event
Reported to Lifeline?
Yes. A silent Maintenance Signal is sent to the Response
Center by the End Device from anywhere wireless signal is
available on the AT&T wireless network. Upon receiving the
Maintenance Signal, the Response Center stores the status
information but no direct action is taken.
Device Hardware Failure
(7000PHB/7000PHW/7000AHB/7000MHB/7100MHB/7200C/7000L)
Alarm Event
Description
In the event that a device senses it has a Hardware Error, a
Warning Message is displayed.
Priority Medium
Device Action May have limited operation, depending on the failure
Alarm Event
Reported/Displayed
Locally?
Yes. Message Button Flashes. Voice Message plays upon
pressing the Message Button.
Alarm Event
Reported to Lifeline?
Yes. A silent Maintenance Signal is sent to the Response
Center by the Communicator. Upon receiving the
Maintenance Signal, the Response Center prioritizes the
response and contacts the Subscriber to verify the status of
their equipment.