Use Instructions

61
Questions? Call 1-800-635-6156
When the device detects a Medium priority alarm: The light on
the GoSafe Mobile Button ashes orange, the device may play
a sound and/or a recorded voice message.
Silencing Alarms
Once a High Priority alarm is detected, it cannot be silenced.
Medium Priority alarms play a sound and possibly a recorded
voice message once upon detection of the alarm, but the light
ashes until the alarm condition is corrected.
Resetting an Alarm
High Priority alarms are normally reset by the Response
Center after the alarm has been responded to. Additionally,
the GoSafe Mobile Button may periodically poll the Response
Center to see if the alarm should be reset and does so
accordingly. If the alarm is not reset by the Response Center,
the GoSafe Mobile Button will automatically reset the alarm
after 30 minutes.
Note: Putting the GoSafe Mobile Button to sleep during an active
high priority alarm will reset the device, but a Voice Message is
played requiring the user to conrm this action. Since the alarm
has already been reported to the Response Center, they will
likely attempt to contact the Subscriber. If an alarm is accidentally
initiated by the User, they should not put the GoSafe Mobile
Button to sleep. Instead, they should wait for the Response Center
to establish voice communication and simply state that the alarm
was sent accidentally.
The GoSafe Mobile Button self-cancels certain Medium
Priority alarms if the cause of the alarm is corrected, stopping
the orange light from ashing.
Agency Submission File