User Manual

For assistance, call 1-800-635-6156
Level
I
m
munity Test
Conducted RF
EN/IEC
61000-4-6
Radiated RF
EN/IEC
61000-4-3
EN/IEC 60601
Test Level
3 Vrms
150 kHz
to 80 MHz
1
0
V/m
80 MHz
to 2.5 GHz
10V/m
Compliance
Electromagnetic
3 Vrms*
Table 4 - Guidance and Manufacturer’s declaration – Immunity
The GoSafe Lite 7150MHB is intended for use in the electromagnetic
environment specified below. The customer or user of the GoSafe Lite
should ensure that it is used in such an environment.
Environment – Guidance
D=1.17√P
D=0.35√P
80 to 800 MHz
D=0.70√P
800 MHz to 2.5 GHz
Por
table and mobile communications
e
quipment should be separated f
rom
the
GoSafe Lite 7150MHB
by no less
than the distances calculated/
l
isted
b
elow
*
:
D=(
3
.
5
/
V1)(√
P)
D=(3.5/E1)(√P)
80 to 800 MHz
D=(7/Eq)(√P)
800 MHz to 2.5 GHz
where P is the maximum power in
watts (W) and D is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,(a) should
be less than the compliance level in
each frequency range.(b)
Interference may occur in the vicinity
of equipment marked with the
following symbol:
* NOTE: Conducted RF Immunity does not apply to an internal battery operated
devices. Limit above is set for battery charger
Confidential Draft for Agency Submission Only.
Not for Posting on any Public-Facing Website
Not for dissemination to the public without the express written consent of Philips.