Device Error (Medium Priority) Alarms
The GoSafe Lite Pendant alarms described below are automatically generated
alarms designed to periodically let the Call Center know that the Pendant is working
properly They are part of normal operation and are the result of no intensional action
by the user.
Alarm Event Type
Low Battery Capacity – 7150 MHB
Alarm Event
Description
When the capacity / life of the battery in the Pendant has
diminished materially and should be replaced, a Warning
Message is displayed
Priority Medium
Device Action Charge cycle frequency increases, otherwise. Operates
normally until battery is depleted.
Alarm Event
Reported/Displayed
Locally?
No.
Alarm Event
Reported to Philips
Lifeline?
Yes. A silent Maintenance Signal is sent to the Response
Center by the device. Upon receiving the Maintenance
Signal, the Response Center prioritizes the response and
contacts the Subscriber to verify the status of
their equipment.
Battery Requires Charging – 7150MHB
Alarm Event
Description
When the rechargeable battery in the Pendant needs to be
charged, a visual and audible warning Message is activated
on the device.
Priority Medium
Device Action Operates normally until the battery is critically low.
Alarm Event
Reported/
Displayed Locally?
Yes. The Indicator Light ashes orange and plays a single
chime. This alarm will self cancel once the battery charge
reaches an acceptable level.
Alarm Event
Reported to Philips
Lifeline?
No.