User Manual

Emissions Test Compliance
RF Emissions
CISPR 11
Group 2
RF Emissions
CISPR 11
Class B
Harmonics
IEC 61000-3-2
N/A for EUT
Class A for charger
Electromagnetic Environment – Guidance
The HomeSafe/GoSafe devices must emit
Electromagnetic energy in order to perform its
intended function. Nearby electronic equipment
may be affected.
The EUTsBattery Powered.
The GoSafe 7000AHB and 7100MHB power
chargers are
suitable for use in all
establishments, including domestic, and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Flicker
IEC 61000-3-3
Table 2 - Guidance and Manufacturer’s declaration – Immunity
The HomeSafe/GoSafe devices are intended for use in the
electromagnetic environment specified below. The customer or user
of the HomeSafe/GoSafe should ensure that it is used in such an
environment.
Complies
EN/IEC 60601
Test Level
Compliance
Level
Immunity Test
Electrostatic
Discharge (ESD)
EN/IEC 61000-4-2
±2 kV for
power
supply lines
±2 kV for
power supply
lines ±1 kV for
input/output
lines
6k
V C
ontact
8kV Air
6k
V C
ontact
8kV Air
Electromagnetic
Environment – Guidance
Floors should be wood, concrete
or ceramic tile. If floors are
synthetic the relative humidity
should be at least 30%
Electrical fast
transient/burst
EN/IEC 61000-4-4
(only for devices with
battery charger power
supply-7000MHB &
7100MHB)
EMC Informational Tables
Table 1 - Guidance and Manufacturer’s Declaration – Emissions The EUT
"HomeSafe/GoSafe devices are intended for use in the electromagnetic
environment specified below. The 7000MHB and 7100MHB devices are
provided with battery power adapter charger. The customer or user of the
HomeSafe/GoSafe should ensure that it is used in such an environment.
Electrical power quality should be
that of a typical commercial or
hospital environment.
Confidential Draft for Agency Submission only. Not for Posting on any Public-Facing website
Not for dissemination to the public without the express written consent of Philips.