Other electronic products exceeding the limits defined in such EMC standards could, under unusual
circumstances, affect the operation of the product.
Medical electrical products need special precautions regarding EMC, and need to be installed and
put into service according to EMC information provided in the accompanying documents.
The use of accessories and cables other than those specified may result in increased emission or
decreased immunity levels.
The product should not be used adjacent to or stacked with other products and that if adjacent
or stacked use is necessary, it should be observed to verify normal operation.
Other equipment could interfere with the medical device or device system, even if the other
equipment complies with CISPR emission requirements.
Caution
Portable and Mobile Phones: Portable and mobile RF communications can affect the HomeSafe /
GoSafe devices. Use caution when using such communication devices within the specified range of
the devices.
Wireless communications equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations, and walkie-talkies can affect this equipment and
should be kept at least a distance 3.3 meters or 11 feet from the HomeSafe/GoSafe devices
(based on a typical cell phone with a maximum output power of 2 W).