IEC regulations
The HomeSafe Help Buttons and Home Communicator are medical electrical (ME)
equipment.
Medical electrical equipment can either generate or receive electromagnetic
interference. This product has been evaluated for electromagnetic compatibility (EMC)
with the appropriate accessories according to IEC 60601-1-2:2007, the international
standard for EMC for medical electrical equipment.
The HomeSafe Help Buttons and Home Communicator must emit electromagnetic
energy in order to perform its intended function. Nearby electronic equipment may
be affected. Medical Electrical Equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided
in the accompanying documents.
The use of accessories, transducers and/or cables other than those specied, with
the exception of those sold by the manufacturer as replacement parts for internal
components, may result in increased emissions or decreased immunity of the
equipment or system.
The equipment or system should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, the equipment or system
should be observed to verify normal operation in the conguration in which it will be
used.
System classication
The HomeSafe Help Buttons and Home Communicator are FDA Class II devices. They
are an internally powered devices for continuous operation.
Radio specications
The HomeSafe Help Buttons and Home Communicator have radios with the following
characteristics: 3 channel frequency agility (917 MHz, 919 MHz, 921 MHz); maximum
EIRP -1.23 dBm; FSK digital modulation. It fully complies with FCC Part 15, Section