8.5. RF Exposure Notice
This Medical Grade Onsight 2500 device intentionally transmits RF energy at 2.4GHz
and 5GHz at up to 65mW for communications. It is recommended this equipment be
kept at a minimum distance of 60cm from all other Medical Electrical Equipment.
See user manual for details of how to turn RF transmit off for non-wireless
configurations.
IEC 60601-1-2:2004(Ed 2.1) Table 202 Requirements
The equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the equipment should assure that it is used in
such an environment.
Emissions test Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
When this device is operating in non-wireless
mode the equipment uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The equipment is suitable for use in all
establishments including domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.