User Manual

25
Leica EG1150 H – Paraffin Embedding Station
The heaters become active and the LEDs indicate the values last set.
The target value LEDs for the heating areas (1, 3, and 4 in Fig. 18) flash
quickly for the duration of the heating phase (cf. "Display of the heating
intervals").
The LED for the paraffin reservoir (2 in Fig. 18) flashes in 1-second inter-
vals and thereby indicates the intensified heating of the paraffin reser-
voir (for 4 hours).
If temperature settings are modified during the heating phase, the in-
strument will store the new settings and the heating phase will then
correspondingly take longer.
Fill the paraffin reservoir with paraffin.
5.2 Switching the instrument on
Activate the standby switch (left-hand side, next to control panel)
(see Fig. 15, 17).
• AllLEDsoftheoperatingpanellightupbrieflyandtheinstalledsoftware
version is shown on the display for around 2 seconds.
All displays go out, the unit changes to standby mode.
The green LED in the ON/OFF button indicates that the instrument is
ready to operate.
5. Operation
Fig. 15
• PresstheON/OFF button down for approx. 2 seconds to
switch over to the operating mode. The time indicator flashes
on the display, indicating that the instrument has been dis-
connected from the power supply.
Press any key to acknowledge.
Fig. 16
Prior to leaving the factory, the Leica EG1150 H is tested thoroughly under laboratory condi-
tions. For this reason, you will find a small quantity of clean, hardened paraffin. You can work
with this paraffin without any problem.
The unit's normal operating mode is STANDBY/ON; i.e. switch on/
off with the ON/OFF key only.
Use the standby switch only if the device is to be switched off for
an extended period.