User Manual
9
Leica EG115 0 C
2.3 Integrated safety devices
Automatic circuit breaker in the main switch
• Anautomatic circuit breaker with alit
switch is located in the main switch. This
circuit breaker disconnects the unit from the
power supply in the event of a short circuit.
• Inthiscase,themainswitchjumpstothe
position "0" = Off.
Fig. 2
Short-circuit
2. Safety
• Theminimumvoltage(seeTechnicaldata)mustbemaintainedwhilestartingtherefrigera-
tion unit.
The compressor needs a start-up current of approx. 25 A. A stable power supply in accor-
dance with the instrument's specifications is essential to its proper functioning. Please
ensure that your electrical installation fulfills these preconditions prior to installing the unit.
Nonobservance causes damage to the instrument.
• Switchofftheinstrumenteachtimebeforeservicing,repairingorcleaning,andpulloutthe
power plug.