User Manual
12
Instructions for Use V 2.4 RevD – 10/2012
4. Instrument Setup
4.2 Delivery range
The basic equipment for the Leica EG1150 C cold plate contains the
following components:
Order No.
• 1basicunitLeicaEG1150C
• 1powercable seep. 15
only for Australia:
• 1powercableEU 14041113558
• 1seriestransformer,230/240V 14046931998
• 1replacementfuseT1.0A,6.3x 32mm 14694301001
• 1InstructionsforUseGerman/English(+CD) 14038880002
• Optional:Unitcover(acrylic)
4.1 Location conditions
• Stable,vibration-freelaboratorytablewithhorizontal,flattabletop,as
far as possible vibration-free ground.
• Nodirectsunlightorstrongtemperaturefluctuations.Roomtempera-
ture consistently between +15 °C and +28 °C.
• Relativeairhumiditymaximum80%,non-condensing.
• Theinstrumentshouldbesetupinsuchawaythattheaircirculation
is not impaired.
At a room temperature of > +28 °C, the working temperature of the
cold plate of -5 °C may not be reached at all points.
To ensure proper function, there must be gap of at least 15 cm
behind the instrument. Failure to observe this distance may result
in serious damage to the refrigeration unit of the device. The instru-
ment should not be operated in hazardous locations.
Please compare the delivered components against the packing list
and your order.
Should there be any discrepancy, please contact the Leica distributor
handling your order.