3.2 Technical data
General data
Approvals: The instrument-specific marks are located on the rear
panel of the instrument next to the name plate
The 120 V version has the test symbol c-CSA-us.
Voltage of power supply:
The device requires a specific AC volt-
age, see table at right.
Instrument type acc. to identification label.
AC ±10 %
Power fuse: T1A L250 V
Nominal power: 350 VA
Maximum start-up current (5 s) 25 A
Operating temperature range: +15 °C to +28 °C
Operating temperatures: -5 °C up to a room temperature of approx. +28 °C
Relative humidity: 20 to 80 % - non-condensing
Permissible temperature range
during storage and transport:
-29 °C to +50 °C
Permissible humidity range
during storage and transport:
10 to 85 % - non-condensing
IEC 1010 classification: Protection class 1
Pollution degree 2
Operating elevation: up to max. 2000 m NN
IP protection class (IEC 60529): IP20
Refrigeration unit
Refrigeration capacity*: 110 W
Safety factor: 3
Refrigerant: R 134a
Compressor oil: 180 cm
3
Emkarate RL 15 s, ICI
*according to CECOMAF, fluid temperature: 55 °C, evaporation temperature: -25 °C
Dimensions and weight
Width: 360 mm
Depth: 650 mm
Height: 360 mm
Weight: 23 kg
Instrument type Power supply voltage Coolant
EG1150 C-1 230 V / 50 Hz 80 g ± 5 g
EG1150 C-3 120 V / 60 Hz 80 g ± 5 g
EG1150 C-8 240 V / 50 Hz 80 g ± 5 g
EG1150 C-11 100 V / 50 - 60 Hz 65 g ± 5 g
3. Instrument Components and Specifications