3. Installation
5
7
Connecting the power cord
• Beforeconnectingthepowercord,ensurethat
the mains switch (1) to Off = 0.
• Fromthecableset,select the power cable
with the plug suitable for your power socket.
Fig. 7
• Insertasmallscrewdriverintothecutout(5)
and carefully pry out the insert.
• Removethefusehousing(8) along with the
fuses (7).
• Replacethedefectivefuses,reinsertthefuse
housing in the instrument and press down until
it snaps into place (with an audible click).
8
1
Before connecting the instrument to
the mains, ensure that the voltage se-
lector switch is set to the correct local
voltage!
An incorrect setting of the voltage se-
lector can cause serious damages to
the instrument!
The instrument must be disconnected
from mains before changing the volt-
age selection.
Connecting to mains
Replacing fuses
Switch the instrument off and unplug
it. Use only the supplied replacement
fuses.
The instrument must be connected to
mains only with the supplied power
cord and only to a grounded socket.
• Connect the plug ofthepowercord to the
connecting socket to the connection socket
(6) on the rear of the control unit.
• Connect the plug ofthepowercord to the
power socket.