User Manual
12
Instructions for Use V1.3 RevC – 11/2012
3. Installation
Fig. 6
3
The instrument MUST be connected to a grounded mains socket.
Only the power cord intended for the local power supply (socket) may be used.
Do not use an extension cord!
The Leica EG F can be connected to different electrical power systems (voltage and frequency-
dependent, refer to Chapter 4.3 "Technical data").
Please observe the following notes to prevent damage to the instrument:
3.4 Electrical connection
The instrument can be connected to the following voltages:
100to120Vand220to240V
~
,
50/60Hz.
Theinstrumentisfactory-setto220to240V
~
(Fig. 6). If the voltage needs to be changed, insert
a screwdriver in the slot (3) and move the setting to the desired value.