3. Installation
3.3 Setting up the instrument/Site requirements
3.2 Standard delivery
Carefully check the delivery against the packing list, delivery note and your order.
Should there be any discrepancy, please contact the Leica sales unit handling your order or
your Leica dealer.
Base unit...............................................................................................................................14 0388 35824
• Controlunitwithvoltageselectorswitch,100-120/220-240V/50-60Hz
• 1screw-ontweezerholderforinstallationonthecontrolunit
• 1screw-inheatableforcepswithspiralcable
• 1setoffuses(2xT1.25AL250V)
• 1setofpowercables
- 1 power cable for Germany ......................................................................................14 0411 13558
- 1 power cable for UK ST/BU F-5A ............................................................................14 0411 27822
- 1 power cable for USA/Canada/Japan ...................................................................14 0411 13559
• UserManual ..................................................................................................................14 0388 83101
• LanguageCD ..................................................................................................................14 0388 83200
• Theinstrumentrequiresaninstallationareaofapprox.150x200mm.
• Thesubstratemusthaveasufficientloadcapacityandrigiditywithrespecttotheweightof
the instrument.
• Relativehumidity:20to80%-non-condensing.
• Roomtemperatureconsistentlybetween+15°Cand+40°C.
• Elevation:uptoamax.2000mabovesealevel.
• Theinstrumentisdesignedforindooruseonly.
• Avoidvibrations,directsunlightandheavyvariationintemperature!
• Thepowersupplymustbewithintherangeofthelengthofthepowercable.Noextension
cable may be connected.
• TheinstrumentMUST be connected to a grounded power socket.
Use only one of the provided power cables that is intended for the local power supply.
Do not operate the instrument in rooms with explosion hazard.
To ensure proper function of the instrument, it must be set up while maintaining a minimum
distance of 10 cm from walls and furniture.