Leica ASP6025
Reagent warning messages
If the RMS is active, warning messages are output if reagents are used for too
long.
The following indicates that the warning threshold values have been exceeded:
• ExclamationmarkonthestationsystemorRTU-bottleintheSMARTscreen
(Fig. 135).
The RMS system – adjusting warning threshold values (continued)
• Whenawarningthresholdvalueisexceeded,awarningmessageisalso
displayed at the end of a program (after the cleaning) (Fig. 137).
From there, you can press YES to go directly to the REAGENT STATUS window.
Fig. 135
Fig. 136
Fig. 137
Fig. 138
• Awarningmessagewhenstartingaprotocol(Fig. 136).
The operator is notied that the reagent has run out and can replace this
reagent before the program starts.
However, it is also possible to start the program despite this.
• Thereagentstatushighlightstheexceededvaluesinredinthetable(Fig. 138).
Displayed if:
• Thenumberofprocessedcassettesisexceeded.
• Thenumberofspeciedcyclesisexceeded.
• Thedaysuntilchangingthereagentareexceeded.
5. Operation