Manual

95
Leica ASP6025
Changing reagent names or reagent groups
• Highlightthereagentthenameorgroupofwhichyouwishtochange.
• Pressthecorrespondingbuttonsymbolintheheadline.
• Intheentrywindow(orviathekeyboard),enterthenewreagentgroupallocation/thenew
reagent name.
• PressOK to save the new reagent group / reagent name.
Entering / modifying reagent thresholds
If warning thresholds are required for a certain reagent, enter them as follows:
• Highlightthereagenttobemodified,eitherbypressingthereagentnameorusingthe
UP/DOWN buttons.
• Intheheaderline,touchtheparametertobechanged-thecorrespondingentryscreenappears
(Fig. 132).
• Enteranewthresholdvalueor-ifnowarningisdesired-pressCLEAR to remove the threshold
altogether.
• PressOK to conrm.
Changeable parameters
Fig. 132
Fig. 133
Conguring the reagent list (continued)
5. Operation
Allocation of reagents to the correct reagent group is the basis for compatibility monitoring. Allocation to the
wrong reagent group can lead to reagent cross contamination.
The respective threshold value applies for all reagent stations containing the
same reagent.
For additional information, refer to Chapter 5.2.2
If a reagent is already used in a program, it can neither be renamed nor can it
be allocated to another reagent group!
The corresponding symbols will be disabled (i.e. they will not be surrounded by
a blue border).
If a reagent is renamed, its allocation to a station will be deleted.