5. Operation
5.2 Reagents
5.2.1 Editing the reagent list
Entering reagent names
This is done via the REAGENTS menu option in the SETTINGS menu. You must be logged on at
supervisor access level to proceed.
• Fromthestartscreen,pressSETTINGS to go to the window of the same name and press the
REAGENTS button there.
• TheSET UP REAGENTS AND WARNING THRESHOLDS window opens (Fig. 130).
• Toaddareagent,proceedasfollows:
• PressINSERT to display the keyboard.
• Enterthenewreagentname.
• PressOK to conrm.
• You willautomatically be promptedto select the reagent groupfor the new reagent
(Fig. 131).
• AllocatethenewreagenttotheappropriategroupandpressOK to conrm.
Fig. 130
Select the reagent group
Fig. 131
The Leica ASP6025 is shipped together with a standard reagent list and predened inltration programs.
The reagent list is adapted to these programs. If the operator writes his own programs, new reagents may have
to be entered.