5.1.10 The Smart Screen
Pressing the SMART SCREEN button in the
main menu opens the SMART FUNCTIONS
window (Fig. 124).
In this window, the entire instrument is repre-
sented graphically.
The representation of bottle positions and
colors on the screen corresponds to the geo-
metrical arrangement and color coding of the
bottles in the instrument.
Fig. 124
Retort status line
The status line displays the actual temperature, the actual
pressure, and the current ll level in the retort
Temperature display
The target temperature of the retort is displayed beside the
retort and the target temperature of the paran appears below
the paran baths.
• Systembottles(15) are designated S1-S9.
• RTU-bottles(10) are designated D1-D6.
• Thereagentcontainedinthebottleisidentiedonthebottlesymbol.
• Thedisplayedlevelsinthebottlesindicatetheapproximateleveltowhichthe
bottles are lled.
• Inthedisplay,thelevelinthecondensatebottle(14) is shown proportionally
according to the number of cycles. When the maximum number of cycles is reached,
a message indicating that it needs to be emptied appears.
• Ifabottlecontainsareagentthathasexceededawarningvalue,thisisshown
with an exclamation mark on a yellow background. For this function to take eect,
threshold values must have been set.