The Service functions menu (continued)
• TheASP6025softwarecontinuouslyrecordswhichreagentwasthelast
reagent in the retort. The last reagent in the retort is displayed next to the
button. This way, it is ensured that incompatible reagents do not come into
contact with one another.
• Ifallingprocesswascanceled,itispossiblethattworeagentsaredisplayed
here.
• Iftheuserknowsthatthereactuallywasadierentreagentintheretortthan
the one that is displayed, this can be corrected by selecting the LAST RETORT
REAGENT button.
• Afterthebuttonhasbeenpressed,rstawarningmessageappearswhich
must be conrmed. Then, the current reagent list is displayed and it is thus
possible to change the reagent status of the retort by selecting the correct
reagent.
Reagent status of the retort
Xylene
The status of the PDF printer is displayed here. It cannot be changed.
"Printer ready" means that a PDF file can be generated and saved after a
program has run, if this was specied in the run options (see also Chapter 5.1.2,
Fig. 94).
Display of printer status
Fig. 107
Printer
ready
5. Operation
Fig. 108
Caution!
The reagent status of the retort may not be changed unless it
is known with certainty that the displayed reagent is not cor-
rect. If incompatible reagents come into contact with one an-
other, this can result in a malfunction of the instrument or the
impairment of a process.