Manual
79
Leica ASP6025
Fig. 104
Active carbon lter
The System monitor menu (continued)
Also in Supervisor mode, only the runtime of the activated charcoal
lter can be reset in the SYSTEM MONITOR menu (Figs. 103/104).
If the ACTIVE CARBON FILTER button is touched, a field appears
which displays the programmed warning and alarm threshold values.
Even in supervisor mode, neither of these values can be changed.
• Ifthestoredwarningvalueisexceeded,awarningmessage
regarding the age of the carbon lter is output.
This warning serves as a reminder to replace the carbon lter. You
can continue using the instrument without any limitations.
• Ifthesetalarmthresholdvalueisexceeded,amessageisoutput
on the screen to inform the user that the charcoal lter age has
been exceeded.
• Whenthismessageappears,atthelatest,theactivecarbonlter
must be replaced in order to ensure that it functions and works
eciently.
5. Operation