Manual
76
Instructions for Use V 2.2 RevI – 08/2013
The access rights to the instrument are congured in this eld in
the form of user proles.
There are three access levels:
• Operatormode
• Supervisormode
• Servicemode
Fig. 100
5.1.3 User proles
SECURITY
The OPERATOR and SUPERVISOR modes have different levels of access and are used to
operate the instrument.
Service mode is intended for Leica service personnel only. A special password is required for
this mode.
The OPERATOR symbol is displayed in the top right corner of the screen.
A password entry is not necessary to switch to operator mode.
Operators can run the following functions:
• Startprograms,enterthenumberofcassettesandchangetherunoptions.
• Interruptandresumerunningprograms.
• Conrmautomaticethanolandxylenereplacement.
• Conrmthemeltingparanandparanreplacementfunctions.
• Runcleaningprograms(exceptsmartcleaning).
• Fillandemptytheretortaswellasallstations.
• Markstationsas"Full" or "Empty" in reagent status.
• EdittheREAGENT GROUPS menu.
• Callup,sortandupdatethesystemstatusandtherunlog.
• Queryresults,printoutPDFles,andsavethemonaUSBstick(see Section 5.1.9).
"Operator" symbol
Operator access level
5. Operation